CETATEST 715 Flow Tester
With intuitive menu control, parameter memory for 64 user-defined test programs, and safety/handling features like password-protected user levels, zero-point adjustment, and overpressure monitoring, the CETATEST 715 enhances traceability and repeatability. Its compressed air supply requirements are modest and interfaces include standard I/O plus optional bus/Ethernet.
Features
Supports −1 bar up to +20 bar (or more by request) gauge pressure for flexible testing of large leakages.
24-bit A/D converter, resolution down to 1 Pa, with multiple measurement units (Pa, ml/min, l/h, etc.).
Pressure decay, sealed part modes, dynamic pressure; programmable phases like prefilling, measuring, venting; 64 test programs memory.
Password protection, zero-result handling, supply pressure monitoring, intuitive menu control.
Delivered with DAkkS calibration certificate (ISO/IEC 17025), 3-year warranty (with yearly maintenance), optional extension.
Overview
| Specification | Detail / Value |
|---|---|
| Operating Principle | Fully automatic leak detection using gauge pressure sensor. |
| Signal Processing | Fast 24-bit A/D-Converter; real-time measurement signal processing by 16-bit µC 40 MHz. |
| Pressure Ranges | Typical options: −1 bar; 200 mbar; 1 bar; 6 bar; 10 bar; 16 bar; 20 bar; negative & positive gauge; custom ranges on request. |
| Resulting Units | Pa, hPa, PSI, Pa/s, hPa/s, mbar*l/s, ml/min, ml/h, l/min, l/h, mmHg, mmWs, Torr. |
| Resolution | 1 Pa (lower requirement ranges). |
| Accuracy / Max Deviation | Gauge pressure: ± 1 % of full scale; Pressure loss test: ± 2 % of full scale. |
| Test Modes | Pressure decay; optional dynamic pressure; sealed component testing. |
| Program Memory | 64 parameterizable test programs, names alphanumeric, export/import via USB or interfaces. |
| Interfaces / Outputs | Standard: RS-232, Digital I/O; optional: Ethernet, Profibus, Profinet; front USB host. |
| Power Requirements | 100-240 VAC, 47-63 Hz; approx. 0.3-0.6 A depending on configuration; max 60 W. |
| Compressed Air Supply | 6-10 bar supply; at least 0.5 bar above test pressure required; under evacuation, 50 mbar under needed. |
| Pneumatic Ports / Connections | Compressed air input 6 mm plug-in fitting; test part port 6 ×1 mm; optional clamp rings, up to 3 pneumatic outputs. |
| Dimensions | 345 × 145 × 435 mm (3U panel size) |
| Weight | Approx. 10 kg |
| Warranty | 3 years with yearly maintenance; optional extension to 5 years. |